A-4 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Technical specifications and default settings: Specifications
NIBP
WARNING
Use only GE CRITIKON blood pressure cuffs. The size, shape,
and bladder characteristics can affect the performance of the
instrument. Inaccurate readings may occur unless GE
CRITIKON blood pressure cuffs are used.
NIBP specifications
Cuff pressure range
(Normal operating range)
0 to 290 mmHg (adult/ped)
0 to 145 mmHg (neonate)
Blood pressure accuracy (Classic and Auscultatory) Meets ANSI/AAMI standard SP-10 (mean error 5 mmHg,
standard deviation 8 mmHg)
Blood pressure accuracy (SuperSTAT) Meets ANSI/AAMI standard SP-10 (mean error 5 mmHg,
standard deviation 8 mmHg)
Maximum determination time 120 s (adult/ped)
85 s (neonate)
Overpressure cutoff 300 to 330 mmHg (adult/ped)
150 to 165 mmHg (neonate)
BP range (Classic and Auscultatory)
Systolic 30 to 245 mmHg (adult/ped)
40 to 140 mmHg (neonate)
MAP 15 to 215 mmHg (adult/ped)
30 to 115 mmHg (neonate)
Diastolic 10 to 195 mmHg (adult/ped)
20 to 100 mmHg (neonate)
BP range (SuperSTAT)
Systolic 30 to 290 mmHg (adult/ped)
30 to 140 mmHg (neonate)
MAP 20 to 260 mmHg (adult/ped)
20 to 125 mmHg (neonate)
Diastolic 10 to 220 mmHg (adult/ped)
10 to 110 mmHg (neonate)
Pulse rate range (Classic and Auscultatory)
30 to 200 beats/min (adult/ped)
30 to 220 beats/min (neonate)
Pulse rate range (SuperSTAT)
30 to 240 beats/min (adult/ped)
30 to 240 beats/min (neonate)
Pulse rate accuracy ± 3.5% or 3 bpm, whichever is higher