Service manual
A-2 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Technical specifications and default settings: Specifications
Specifications
General
General specifications
Mechanical
Dimensions
Height 19.5 cm (7.7 in)
Width 21.9 cm (8.6 in) without temperature
25.4 cm (10.0 in) with temperature
Depth 13.5 cm (5.3 in)
Weight (Including battery) 2.4 kg (5.4 lb)
Mountings Self-supporting on rubber feet, pole mounted, or wall mount bracket
Portability Carried by recessed handle
Power requirements
Power converter universal P/N: 2018859-001
Protection against electrical shock Class II
AC input 100 to 240VAC, 0.5A
DC output voltage 12VDC at 1.0A
The AC mains power adapter contains a nonresettable and
nonreplaceable fuse.
Monitor
Protection against electrical shock Internally powered or Class II when powered from specified external
power supply.
DC input voltage 12 VDC, supplied from a source conforming to IEC 60601-1.
Fuses The monitor contains three fuses. The fuses are mounted within the
monitor. The fuses protect the low voltage DC input, the main battery,
and the remote alarm output. The +5 V output on the host port connector
is regulated by internal supply.
Main Battery Refer to “Battery” on page A-15
Environmental
Operating temperature + 5°C to + 40°C (+ 41°F to + 104°F)
Operating atmospheric pressure 500 hPa to 1060 hPa