Service manual

6-42 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Parts lists and drawings: Disassembly/reassembly of FRUs
6. Remove sub-chassis and set aside.
7. Remove pneumatic assembly from sub-chassis.
SpO
2
board
1. Remove 2 screws securing SpO
2
board.
Threaded standoffs have screws on both ends.
2. Remove SpO
2
board.
Front bezel
1. Remove 2 SpO
2
connector screws.
2. Unplug 2 pneumatic hoses from bezel.
3. Remove 4 torx-head screws along the bottom of main board.
4. Lift unit and remove bezel.