6-40 CARESCAPE V100 Vital Signs Monitor 2048723-002D
Parts lists and drawings: Disassembly/reassembly of FRUs
3. The battery can now be removed.
Rear case
1. Remove remaining two screws from the bottom of the monitor.
2. Set monitor upright.
3. Carefully remove the front faceplate.
4. Remove 2 screws behind black overlay.
5. Set monitor on its face and open the printer door.
6. Carefully remove the rear case.