Service manual

2048723-002D CARESCAPE V100 Vital Signs Monitor 6-39
Parts lists and drawings: Disassembly/reassembly of FRUs
8. Place the tabs on the bottom edge of the face plate into the slots on the
monitor.
9. Snap the face plate back on the monitor.
Monitor disassembly procedure
The following procedure is sequential (i.e., you must remove the main battery
and rear case to remove the printer, etc.).
CAUTION
Internal electronic components are susceptible to damage by
electrostatic discharge. To avoid damage when disassembling
the monitor, observe the standard precautions and procedures
for handling static-sensitive components.
Main battery
1. Remove 4 screws securing the instruction cards.
2. Remove the battery compartment door and adhesive pad.