and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL
MAINTENANCE
Continued
CALIBRATION
BLOOD PRESSURE
CALIBRATION CHECK
Replacement of Battery Fuse
Refer to qualified service personnel since replacement of the battery fuse
requires unit disassembly.
Calibration of the monitor should be checked at least once every three
months or when there is doubt about the validity of the pressure readings.
CAUTION
Calibration equipment should always be kept dry and free of
particulate matter. Moisture or foreign substances introduced
into the pneumatic system can cause damage to the unit.
MANOMETER
NOTE: Use Adult cuff
(Reorder No. 2774) and
DINAMAPT”
PORTABLE