PORTABLE ADULT/PEDIATRIC and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL
MAINTENANCE
Continued
Figure 6-5.
Open AC Line Fuse Compartment
Using the same tool, pry out each fuse holder as shown in Figure 6-6.
Figure 6-6. Remove Fuse Holder
Remove each fuse from its fuse holder as shown in Figure 6-7 and inspect it
for a burned or broken filament. If the filament appears to be intact, check it
for continuity with an ohmmeter. Replace the blown fuse with one of the
exact same rating and type.
42