ambient temperature.
Battery Removal and Replacement
Under normal conditions the battery should last at least two years. It is best
to keep the battery charged as fully as is practical and never store the
monitor with the battery in a discharged condition. When the battery will no
longer take a charge, remove it and replace it with one of the same part
number as outlined below.
1.
Ensure that AC power is removed from the monitor either by removing
the power cord or placing the AC power switch on the rear of the unit to
the OFF (0) position.
2.
Using a dime, or similar tool, loosen the battery door hold-down screws
as shown in Figure 6-1. Turn each screw counterclockwise at least
one-half turn. Do not attempt to remove screws.
38
ADULTIPEDIATRIC
and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL
MAINTENANCE
Continued
TABLE 6-1. BATTERY CHARGING CHARACTERISTICS
CHARGE TIME*
OPERATING TIME**
30
minutes
15
minutes
1
hour
30
minutes
2
hours
1 .O
hour
3
hours
1.6
hours
4
hours
2.2 hours
5
hours
2.7
hours
6
hours
3.2 hours
7
hours
3.7 hours
8
hours
4.1
hours
10
hours
4.6 hours
16
hours
5.3 hours
24 hours
6.0 hours
*Charging at 25°C ambient temperature.
**Operating with standard adult cuff on 5 minute determination cycles at
DINAMAPTM
PORTABLE