Press AUTOIMANUAL to enter auto mode and start a determination.
Auto mode is indicated by the LED above the AUTOIMANUAL switch
and the appearance of the cycle time in the CYCLE MINUTES display
28
mmHg)
Momentarily press the SELECT or CANCEL switch to return the display
to normal operation.
To operate the monitor in manual mode, follow this procedure.
1.
Press START to begin each determination.
2.
Press CANCEL to stop a determination and deflate cuff, or to cancel
overpressure, excess time and patient alarm indicators.
mmHg)
Momentarily press the LOW LIMIT switch to display the current low
limit in the DIASTOLIC display. Press and hold LOW LIMIT to cause the
monitor to increment through the low alarm range settings. Release
the LOW LIMIT switch at desired setting. (Range: 15 to 120
mmHg)
Momentarily press SELECT to enable the DIASTOLIC display to show
the diastolic alarm limits, All other displays will be blanked and the
DIASTOLIC display will show zero (or the diastolic value of the previous
determination).
Momentarily press HIGH LIMIT to display the current diastolic high
limit. Press and hold HIGH LIMIT to cause the monitor to increment
through the high alarm range settings. Release the HIGH LIMIT switch
at the desired setting. (Range: 50 to 180
mmHg)
Momentarily press the LOW LIMIT switch to display the current low
limit in the PULSE display. Press and hold the LOW LIMIT switch to
cause the monitor to increment through the low alarm range settings.
Release the LOW LIMIT switch at the desired setting. (Range: 40 to
140
ADULTIPEDIATRIC
and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL
SET ALARM LIMITS
PROCEDURE
Continued
MANUAL MODE
AUTO MODE
7.
a.
9.
10.
11.
12.
13.
Momentarily press the SELECT switch to enable the PULSE display to
show the pulse alarm limits. All other displays will be blanked and the
pulse display will show zero (or the pulse of the previous determination
if one was performed).
Momentarily press HIGH LIMIT to display the current high limit in the
PULSE display. Press and hold HIGH LIMIT to cause the monitor to
increment through the high alarm range settings. Release the HIGH
LIMIT switch at the desired setting. (Range: 80 to 220
DINAMAPTM
PORTABLE