Momentarily press the LOW LIMIT switch to display the current low
limit in the MAP display. Press and hold the LOW LIMIT switch to
cause the monitor to increment through the low alarm range
settings, Release the LOW LIMIT switch at the desired setting.
(Range: 25 to 120
mmHg)
Momentarily press the SELECT switch to enable the MAP display to
show the MAP alarm limits. All other displays will be blanked and
the MAP display will show zero (or the MAP of the previous
determination if one was performed).
Momentarily press HIGH LIMIT to display the current high limit in the
MAP display. Press and hold HIGH LIMIT to cause the monitor to
increment through the high alarm range settings. Release the HIGH
LIMIT switch at the desired setting. (Range: 70 to 200
mmHg)
Momentarily press the LOW LIMIT switch to display the current low
limit in the SYSTOLIC display. Press and hold LOW LIMIT to cause
the monitor to increment through the low alarm range settings.
Release the LOW LIMIT switch at desired setting. (Range: 30 to 150
CRITIKON, INC.
SECTION 5. INSTALLATION AND OPERATION
SET ALARM LIMITS
PROCEDURE
The adult default alarm limits are in effect each time power is first
applied to the monitor. If neonatal default alarm limits are desired, make
one determination with a neonatal cuff and hose attached prior to
inspecting or changing limits. See Table 5-2, Default Alarm Limits.
To change these limits, perform the following steps. However, note that if
no switch is pressed within a lo-second period while setting limits, the
monitor automatically returns to the normal display.
1.
2.
3.
4.
5.
6.
Momentarily press SELECT to enable the SYSTOLIC display to show
the systolic alarm limits. All other displays will be blanked and the
SYSTOLIC display will show zero (or the systolic value of the
previous determination).
Momentarily press HIGH LIMIT to display the current systolic high
limit. Press and hold HIGH LIMIT to cause the monitor to increment
through the high alarm range settings. Release the HIGH LIMIT
switch at the desired setting. (Range: 75 to 240