Service manual

receptable
to be used.
UPON INITIALLY RECEIVING THE MONITOR, CHARGE THE BATTERY
FOR SIXTEEN HOURS by connecting the detachable lo-foot power
cord to the monitor rear panel. Plug the other end into an appropriate
voltage receptacle and place the rear panel POWER switch into the ON
(ā€œIā€) position. (See Section 6, Maintenance for battery charging chart).
Connect the pneumatic hose to the monitor at the front panel. There is
no preferred order of connection; either hose connector may be
attached to either port. Thread the hose connectors onto monitor ports
until finger-tight. DO NOT OVERTIGHTEN. The pneumatic seal is not
made by tightening the connector.
22
DINAMAPTM
PORTABLE ADULT/PEDIATRIC and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL
INSTALLATION AND
INITIAL SETUP
Electrical and Hose Connections
If there is no apparent external damage to the monitor, follow these steps
With monitor power off:
Check the voltage rating stamped on the serial number plate attached
to the bottom of the monitor and make sure it matches the line voltage
of the