Service manual

-I
Figure 4-2.
Basic Operating Cycle
17
-
auto and manual. Wait time is affected by the
cycle time (auto mode) or operator intervention (manual mode). See
figure 4-2.
I
-CYCLE TIME
0
and wait time.
Inflation time, deflation time, and evaluation time are the same for both
modes of operation
0
evaluation time,
0
deflation time,
0
inflation time,
170mmHg
Max 225mmHg
Arterial
Min 20mmHg
Min
20mmHg
Pressure)
Pulse
Max 220bpm
Max 200bpm
Rate
Min 40bpm
Min 40bpm
The operating cycle is comprised of four parts:
16OmmHg
Max 21 OmmHg
Pressure
Min
1 OmmHg
Min
1 OmmHg
MAP (Mean
Max
3OmmHg
Diastolic
Max
19OmmHg
Max 245mmHg
Pressure
Min 30mmHg
Min
CRITIKON, INC.
SECTION 4. FUNCTIONAL DESCRIPTION
MAXIMUM-MINIMUM
RANGES
BASIC OPERATING
CYCLE
Maximum and minimum ranges of Blood Pressure and Pulse Rate
determinations are shown in Table 4-l.
TABLE 4-1.
MAXIMUM AND MINIMUM DETECTABLE PRESSURE
Parameter
Neonatal
Adult/Pediatric
Systolic
Max