Service manual
0
allows the HIGH LIMIT and LOW LIMIT
switches to be used to display (set) the high
and low pulse rate alarm limits in the PULSE
display.
12
0
blanks all displays except PULSE, and
0
allows the HIGH LIMIT and LOW LIMIT
switches to be used to display (set) the high
and low MAP alarm limits in the MAP display.
Pressing the switch a third time
0
allows the HIGH LIMIT and LOW LIMIT
switches to be used to display (set) the high
and low systolic alarm limits in the SYSTOLIC
display.
Pressing the switch a second time
q
blanks all displays except MAP, and
0
blanks all displays except SYSTOLIC, and
entrre
range of high limit settings (settings overlapping
the low limit will not be displayed). The new high
limit alarm setting will be the setting that is
displayed at the time you release the HIGH LIMIT
switch.
This grey momentary pushbutton switch selects
the alarm limits to be displayed.
Depressing the switch the first time
ia
LOW
LIMIT
SILENCE
HIGH
LIMIT
SELECT
This grey momentary pushbutton switch displays
the current low alarm limit that you select (SELECT
switch). If you push and hold this switch, you will
cause the display to step through the entire range
of low limit settings (settings overlapping the high
limit will not be displayed.) The new low limit alarm
setting will be the setting that is displayed at the
time you release the LOW LIMIT switch.
This grey momentary pushbutton switch with LED
indicator alternately mutes and enables the audio
alarm. When pressed (SILENCE ON) the LED
indicator lights to indicate that the audio portion of
the alarms have been silenced. When an alarm
condition is detected the LED will flash but no
audio alarm will sound. Press again to enable
audio alarm.
This grey momentary pushbutton switch displays
the current high alarm limit that you select
(SELECT switch). If you push and hold this switch,
you will cause the display to step through the
TABLE3-1.
Model8100 FRONT PANELCONTROLSAND
INDICATORS Continued
ITEM NAME
FUNCTION
15
16
17
DINAMAP’” PORTABLE ADULT/PEDIATRIC and NEONATAL VITAL SIGNS MONITOR OPERATION MANUAL