Service manual

45,60, and 90 minutes.
Holding the SET switch in for 4 seconds while
momentarily pressing the ON switch places the
monitor in calibrate mode, where it will remain for
3 minutes. Holding the SET switch for 20 seconds
while momentarily pressing the ON switch will cause
the monitor to remain in calibrate mode indefinitely.
15,20,30,
3,4,5,
10, 1,2,
0
shows diastolic alarm limits (see Item 18,
SELECT).
This green momentary pushbutton switch initiates a
determination in manual or auto mode. The START
switch initiates a determination only during wait time;
pressing START switch during a determination has no
effect.
In manual mode, determinations are initiated by
pressing the START switch. In auto mode, pressing
the START switch will initiate a determination and
begin a new cycle.
This red momentary pushbutton switch performs
several functions. Pressing this switch will 1) ter-
minate a determination in progress; 2) cancel all
visual and audible alarms except certain system
alarms (refer to Table 5-3, System Alarms Summary
table); 3) exit the calibrate mode; 4) exit patient alarm
set routine; and 5) exit Prior Data Displayed Mode.
Visual and audible alarms may be cancelled during
wait time and the displays will maintain the last
determined values.
This grey momentary pushbutton switch increments
the MINUTES display when in auto mode. The
MINUTES increments are
3-digit
red LED display shows the
diastolic pressure. In addition this display:
3-digit
red LED display shows systolic
pressure. In addition, this display
q
shows systolic alarm limits (see Item 18, SELECT).
This
El
shows MAP alarm limits (see Item 18,
SELECT).
This
0
flashes cuff pressure during deflation time.
mmHg
START
CANCEL
SET
This 3-digit red LED display shows mean arterial
pressure. In addition this display;
mmHg
DIASTOLIC
mmHg
SYSTOLIC
INDICATORSContinued
ITEM NAME
FUNCTION
MAP
TABLE3-1.
MODEL 8100 FRONT PANEL CONTROLS