Service manual
mmHg
standard deviation.)
PULSE RATE DETERMINATION
Adult/Pediatric
(Maximum) 200 bpm
(Minimum)
40 bpm
Neonate
(Maximum) 220 bpm
(Minimum)
40 bpm
7
mmHg
mean error; 8
&
5
AAMI
standards
for non-invasive blood pressure accuracy. (AAMI standard:
mmHg
BLOOD PRESSURE ACCURACY
Blood pressure accuracy meets and exceeds proposed
mmHg
(Minimum)
10
mmHg
Neonate
(Maximum) 160
mmHg
(Minimum)
10
mmHg
DIASTOLIC DETERMINATION
Adult/Pediatric
(Maximum) 210
mmHg
(Minimum)
20
mmHg
Neonate
(Maximum) 170
mmHg
(Minimum)
20
mmHg
MAP DETERMINATION
Adult/Pediatric
(Maximum) 225
mmHg
(Minimum)
30
mmHg
Neonate
(Maximum) 190
mmHg
(Minimum)
30
mmHg
SYSTOLIC DETERMINATION
Adult/Pediatric
(Maximum) 245
? 15
+ 15mmHg
Neonate
125
mmHg
INITIAL CUFF INFLATION
Adult/Pediatric
178
mmHg
to 235
mmHg
Neonate
0
mmHg
to 250
CRITIKON, INC.
SECTION 3. PHYSICAL DESCRIPTION
SECTION 3. PHYSICAL DESCRIPTION
The DINAMAP’” Portable Monitor Model 8100 is a microprocessor
controlled noninvasive device housed in a blue plastic case.
Performance and technical specifications for the Model 8100 follow.
Tables 3-l and 3-2 contain a general physical description of the
monitor’s controls, indicators, connectors, and operating requirements
PERFORMANCE
SPECIFICATIONS
CUFF PRESSURE RANGE
Adult/Pediatric
0