6-20 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
SpO
2
Board
1. Remove hex nuts near SPO2 board.
2. Turn monitor over and remove 2 SPO2 connector screws.
3. Turn monitor on its face.
4. Using needle nose pliers squeeze nylon retention tabs and remove SPO2
board.
Main Board
1. Remove 5 phillips screws.
2. Remove 2 torques head screws.
3. Unplug tubing from main bd.