Service manual
6-18 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Parts List, Drawings, and Replacement: Assembly/Disassembly of FRUs
3. Carefully remove the front faceplate.
4. Remove 2 screws behind black overlay.
5. Set monitor on its face and open the printer door.
6. Carefully remove the rear case.
7. Unplug speaker cable from circuit board.
8. Unplug rear case cable.
Printer
1. Unplug air hose from valve and 2 elbow joints.
2. Remove 3 cable assemblies.