Service manual

Revision D DINAMAP
®
ProCare Monitor 5-5
2009381-001
Troubleshooting: Alarm Code Interpretation
E13 Battery Battery too low to operate the unit No Charge battery
External DC source failed
Replace battery
E21 SPO2 Failed SPO2 sensor Replace SPO2 sensor
E23 SPO2 SPO2 Sensor off finger Yes Reposition SPO2 sensor
E25 SPO2 SPO2 No signal Yes Reposition SPO2 sensor
E61 Temp Temp probe failure No Replace temperature probe
E63 Temp Probe disconnected or wrong type Yes Check for correct probe
E66 Temp Temperature probe too hot No Replace probe
E80 NIBP Overpressure Yes Check for pinched or occluded
internal tubing
E82 NIBP Excess air in cuff Yes Wrong cuff type
Failed valve on pneumatics
assembly
E83 NIBP NIBP inflation time out Yes Leak in cuff or o-ring in hose
Internal leak in tubing or
pneumatic valve
Pump not turning on
E84 NIBP NIBP total time out Yes Pneumatic assembly failure
E85 NIBP NIBP level time out Yes Cuff placement on patient
Minor leak in cuff
Pneumatic assembly failure
E89 NIBP NIBP no pulse detected Yes Cuff placement on patient
*Acknowledging an alarm by pressing the Silence button, cancels the alarm.
** Call Technical Support for definitions and instructions.
Error Code
or Problem
Source Definition
Can be
Silenced?
Probable Source