Service manual
5-4 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Troubleshooting: Alarm Code Interpretation
8. The SPO2 window shows the Time the error occurred.
9. The Pulse Rate window shows the error code that occurred at the recorded
time.
10. To print the error log, press the Print button while viewing the log.
Error Code
or Problem
Source Definition
Can be
Silenced?
Probable Source
Display
blank, high
pitch alarm
System Failsafe error No Mains PWA issue
930 SPO2 No status from module for 30 ±10 sec.
Fatal error reported by module
No Check cable and sensor
Parameter turned on - no
hardware installed in unit
940 Temp TEMP data samples less than 45 in 5
sec while idle
No Incorrect or failed temperature
probe
950 NIBP NIBP pump on during idle or over
current detected
No Pneumatic assembly failure
951 NIBP NIBP valve stuck closed during cuff
typing
No Pneumatic assembly failure
952 NIBP NIBP PT2 higher than 150 for greater
than 15 seconds while idle
No Pneumatic assembly failure
970 Printer Time base failure No Mains PWA issue
971 System RAM test failure No Mains PWA issue
972 System ROM checksum failure No Mains PWA issue
973 System Secondary SPI communication error
during initialization
No Mains PWA issue
974 System Calibration data invalid on initialization
or unit never calibrated
No Calibrate unit
Mains PWA issue
975 System Calibration data storage exhausted
while saving data in CAL Mode
No Return to depot for software
installation
E00 System Memory lost Yes Usually noted after changing
batteries. User settings and date/
time revert to default settings.
E10 Printer No paper in printer Yes No paper in printer
Printer problem
E11 Printer Printer too hot Yes Printer problem
E12 Battery Battery to low to print Yes Charge battery
Replace battery