Revision D DINAMAP
®
ProCare Monitor 4-17
2009381-001
Maintenance: Test Results Form
Test Results Form
Description Min Max Actual Pass-Fail-N/A
LEAKAGE
Leakage Result (mmHg) 0 6
PRESSURE TRANSDUCER VERIFICATION
Pressure reading at 200mmHg, top display - Systolic 197 203
Pressure reading at 200mmHg, bottom display -
Diastolic
197 203
Pressure reading at 150mmHg, top display - Systolic 147 153
Pressure reading at 150mmHg, bottom display -
Diastolic
147 153
Pressure reading at 100mmHg, top display - Systolic 97 103
Pressure reading at 100mmHg, bottom display -
Diastolic
97 103
Pressure reading at 50mmHg, top display - Systolic 47 53
Pressure reading at 50mmHg, bottom display - Diastolic 47 53
OVERPRESSURE VERIFICATION
Overpressure threshold, Adult (mmHg) 305 325
Overpressure threshold, Neonate (mmHg) 150 165
BUTTONS
BP Determination Initiated
"E80" displayed on SYSTOLIC display
Audible alarm can be silenced
"Silenced" LEDs flash
Overpressure alarm can be cleared
Alarm button is functioning
DISPLAY
All 7-Segment LEDs Light Correct Color
All Discrete LEDs Light, Correct Color
EXTERNAL DC DETECTION
Charging indicator LED illuminated
BP DETERMINATION
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)
Heart rate reading (bpm)
Systolic reading (mmHg)
Diastolic reading (mmHg)
MAP reading (mmHg)