Service manual
4-16 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Maintenance: Parameter Level Functional Testing
SpO
2
(Perform only if equipped with SpO
2
module)
1. Connect the appropriate SpO
2
probe and cable to the SpO
2
connector. Place
the probe on your finger.
2. Verify the unit displays a:
Pulse value
Saturation value
Signal Strength Bar Graph
3. Remove the sensor from your finger to generate an E23 alarm and to sound
the speaker.
4. Press the Silence button.
5. Verify the sound has stopped and the E23 error is cleared.
6. Re-apply the SpO
2
sensor to your finger.
7. Verify the unit displays a:
Pulse Value
Saturation value
Signal Strength bar Graph
Printer Output Test
1. Load Paper into the print mechanism.
2. Press Print button.
3. Verify the printer outputs a record and print quality is good.
Temp Circuit Leakage Test
1. Setup an IEC 60601-1 approved leakage tester to apply 240 VAC to an
isolated circuit.
2. Plug temp probe Hi Pot adapter into the temp jack.
3. Record and verify the temp circuit leakage current.
SpO
2
Circuit Leakage Test
1. Setup an IEC 60601-1 approved leakage tester to apply 240 VAC to an
isolated circuit.
2. Plug an SpO
2
cable into the SpO
2
connector on the front of the unit.
3. Plug SpO
2
probe Hi Pot adapter in to the DB9 jack at the end of the SpO
2
cable.
4. Record and verify the SpO
2
circuit leakage current.