Service manual
4-14 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Maintenance: Parameter Level Functional Testing
External DC Verification
1. Plug the power brick into the Monitor.
2. Verify that the CHARGING indicator is illuminated.
BP Determination
NOTE:For best results, it is important to be seated and not moving, talking,
eating, and/or smoking, etc. while taking your blood pressure. If you are
uncertain as to the proper technique, consult the operation manual.
1. Remove the calibration set up and attach a normal adult cuff and hose (be
sure to select the correct cuff size).
2. Press Start/Stop button on monitor to begin a determination.
3. Record Systolic, Diastolic, map and heart rate from the Monitor display.
4. Wait 1 minute, then press Cycle button to initiate a determination in Auto BP
mode.
5. Record Systolic, Diastolic, map and heart rate from the Monitor display.
6. Wait 1 minute, then press Cycle button until st is displayed in the min
window to initiate a determination in STAT mode.
7. Record Systolic, Diastolic, map and heart rate from the Monitor display.
8. Press Start/Stop button, end STAT mode.
BP Overpressure Verification
1. Remove the cuff/hose from the monitor, restrict airflow from cuff hose port.
2. Press Start/Stop to begin BP determination.
3. Verify E80 is displayed on the Systolic window and audible alarm sounds.
4. Remove the air restriction.
5. Press Start/Stop and verify that the pump does not start.
6. Press the Silence button.
7. Press the Silence button again.
8. Verify the alarm condition is cleared from the Systolic window.
Temperature (Perform if equipped with Temp module)
The Temperature Simulator for the Alaris System is available from Alaris Medical
Systems, Inc. (619) 458-7000.
The Alaris Turbo Temp probes cannot be calibrated. These probes must be
discarded after 2 years from the date of manufacture stamped into the RJ45
connector (first two digits=year, second two digits=week). Refer to the illustration
example (0520=fw20 in 2005).
It is the responsibility of the user to maintain proper records.