Service manual
Revision D DINAMAP
®
ProCare Monitor 4-13
2009381-001
Maintenance: Parameter Level Functional Testing
6. Close valve on bulb and slowly inflate pressure to 200 mmHg (using the
manometer as reference).
7. Press Menu button when pressure reads exactly 200 mmHg to save
calibration setting.
8. Press Cycle button until 6 appears in the min window.
9. Press and hold Menu button until monitor beeps, which acknowledges that
data was saved.
10. Turn the Monitor off.
Overpressure Verification
NOTE:: To enter service mode press and hold Cycle button while pressing the
On/Off button.
1. Wait a few seconds after entering service mode. Press Cycle button so that
the min window changes from 0 to 1.
2. Use the inflation bulb to inflate close to 300 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
3. Record and verify pressure at which valve opens.
4. Press Cycle button and 2 should appear in the min window.
5. Use the inflation bulb to inflate close to 150 mmHg. Slowly inflate (1 to 2
mmHg/sec) until valve opens and pressure is released.
6. Record and verify pressure at which valve opens.
7. Turn unit off.
Button Testing
1. Disconnect the cuff/hose assembly and power on the unit.
2. Press Inflate/Stop button.
3. Verify a BP determination has been initiated.
4. Block pump port and verify E80 alarm.
5. Verify flashing red indicator of Silence button.
6. Press Silence button, verify alarm has been removed.
7. Press Alarm button several times, verify unit cycles through all alarm
settings (i.e., SYS, DIA, SpO
2
).
8. Turn unit off.
LED Tests
1. Power on the Monitor.
2. During the power-up self-test verify all 7 segment LED display segments and
all discrete LEDs (except CHARGING LEDs) illuminate.
3. Repeat power up cycle until all LEDs are checked.