Service manual
4-8 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Maintenance: Long-Term Storage
Replacing the Battery
1. Unplug the Monitor from the DC power source.
2. Looking at the bottom of the ProCare Monitor, remove the battery
compartment cover by removing the four screws that secure the cover and
help card tray.
3. Remove the help card tray and battery door cover.
4. Remove the old battery and disconnect the wires. Attach the battery wires
to the new battery, ensuring the red terminal (+) is connected to the red wire
and the black terminal (-) is connected to the black wire.
5. Insert the battery into the compartment.
6. Then replace the cover, help card tray, and screws. Insert the external DC
power converter plug into the external DC power socket and plug into an AC
outlet.
NOTE: Error code E00 appears (Memory Lost) alerting you that the user
settings (including alarm limits and inflation pressure) and date/
time will go back to factory default values.
7. Reset the date/time and applicable user settings.
CAUTION
Do not touch either the pin of the DC input connector or the
terminals within the battery compartment and the patient at
the same time.