Service manual
Revision D DINAMAP
®
ProCare Monitor 3-13
2009381-001
Installation: Host Communications Connector
Host Communications Connector
All host port signals are NON-ISOLATED and should be connected to equipment
conforming to IEC 60601-1-1 ONLY. Where isolation of data communication is
required, the isolated level converter should be used. If external alarm control is
required, p/n 487208 (Isolated Remote Alarm Cable Assembly) should ALWAYS
be used. Please refer to the Information Sheet included with the isolated remote
alarm cable for operational details.
NOTE: When using remote alarm, the ProCare Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.
DB15 Connector Pin Assignments
Connection Details
Host Port Connector (rear panel)
WARNING! Auxiliary equipment connected to the DINAMAP
®
ProCare Monitor
will result in the formation of an electromedical system and thus, must comply
with the requirements of EN 60601-1-1/ IEC 60601-1. All host port signals are
NON-ISOLATED and should be connected to equipment conforming to IEC
60601-1, configured to comply with IEC 60601-1-1 ONLY. Where isolation of
data communication is required, GE Medical Systems Information Technologies
part number ILC1926 should be used. If external alarm control is required, GE
Medical Systems Information Technologies part number 487208 (Isolated
Remote Alarm Cable. When a high-priority alarm condition is displayed on the
Monitor, the remote alarm signal becomes active within 0.5 seconds. The active
state of the alarm signal is an open circuit. In the inactive state the alarm signal
is connected to ground. Please refer to the Information Sheet included with the
isolated remote alarm cable for operational details.
NOTE: When using remote alarm, the ProCare Monitor should be considered
the primary alarm source. The secondary alarm is used for secondary
purposes only.