Service manual

Revision D DINAMAP
®
ProCare Monitor 3-9
2009381-001
Installation: Configuring Your ProCare Monitor
Procedure for units with Masimo® Technology
(Refer to the Masimo® section for options)
1. Enter the configuration mode: with the Monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds.
2. Press the Menu button until
n0d appears in the Pulse Rate window.
3. Use the +/- buttons to select the option.
4. Press the Menu button once. SAt appears in the Pulse Rate window.
5. Use the +/- buttons to select the option.
6. Press the Menu button once. SEn appears in the Pulse Rate window.
7. Use the +/- buttons to select the option.
8. To exit the configuration mode, turn the unit off. To continue with additional
configuration settings, press Menu.
Advanced Configuration Mode
Advanced Configuration mode is entered by holding the Menu button and the -
button simultaneously while powering on with the On/Off button.
For a couple of seconds immediately after power up in this mode the Systolic
and Diastolic display windows will display the major and minor version codes.
The version codes are numbers that represent the letters of the English alphabet
which are designated to the currently loaded version of the monitor firmware
(e.g., 1 indicates A, 2 indicates B, etc.).
After a moment, this version information is cleared, and the monitor displays the
1
st
page of configuration mode which simply displays ACF in the Systolic display
window indicating that the monitor is in advanced configuration mode.
You can then press the Menu button to cycle through all the advanced
configuration mode option pages. After all options pages have been displayed,
the display will return to the 1
st
advanced configuration mode page (displaying
ACF). You can use the + and - buttons to make changes to settings. After
making changes, simply cycle the power to return to normal operation mode.
Changes are automatically be retained.