Service manual
Revision D DINAMAP
®
ProCare Monitor 3-7
2009381-001
Installation: Configuring Your ProCare Monitor
Pressing the Menu button cycles through all the configuration option pages.
After all options pages have been displayed, the display returns to the 1
st
configuration mode page (displaying CFG). You can use the + and - buttons to
make changes to settings. After making changes, simply cycle the power to
return to normal operation (clinical) mode. Changes are automatically retained.
The Menu selections appear in the following order. Refer the each manual
section for settings options.
NOTE:Menu selections for a Masimo unit have three SpO
2
settings. Refer to the
Masimo SpO
2
section for options.
Setting:
Window LED Display
• Inflate
Pressure: Diastolic XXX (numeric)
• SpO
2
Mode: SpO
2
• SpO
2
Sat: SpO
2
• SpO
2
* Sensitivity: SpO
2
• Temp: °C or °F
• Year: Systolic
• Month: MAP/Cuff
• Day: Diastolic
• Hour: min
• Minute: min
• Mode: Systolic
*Masimo units only
Setting the Date and Time – To set the date and time on the ProCare Monitor,
you must access the configuration mode. Press Menu to skip the default settings
that do not require changes. The following list shows the windows in which the
date/time settings appear.
Setting:
Window LED Display
• Year: Systolic
• Month: MAP/Cuff
• Day: Diastolic
• Hour: min
• Minute: min
Procedures
1. Press the Menu button to move from one setting to another. Use the +/-
buttons to increment or decrement the setting.
NOTE: For the date and time to be saved, you must advance the menu
through the minute setting.
2. To exit the configuration mode, press the On/Off button.
3. To continue with other changes, press the Menu button. CFG will appear in
the Systolic window. To change parameter settings, press the Menu button
and select the parameter function. To change alarm settings, press the
Alarms button.