Service manual
3-6 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Installation: Configuring Your ProCare Monitor
Configuring Your ProCare Monitor
Operating Modes
The ProCare Monitor can operate in one of four modes: clinical, configuration,
advanced configuration, and service.
Clinical Mode
Clinical mode is the Monitor’s normal operating mode. While this mode is active,
alarm limits and a few other commonly used settings are adjustable. All
parameters are available for monitoring in this mode.
Configuration Mode
Configuration and advanced configuration modes display the software revision
and allow you to configure defaults for some settings that are available in
clinical mode, as well as some less commonly used settings that are only
adjustable in these modes. A fatal error history is also available in the advanced
configuration mode. No parameters are operable in these modes, therefore,
patient monitoring should be suspended.
Configuration Mode Settings
Monitor settings such as HIGH/LOW alarm settings changed in the Clinical Mode
will not be retained after the monitor is powered off. To retain alarm and
parameter settings, the changes must be done in the configuration mode. Date/
Time settings are also entered in the configuration mode.
To enter the configuration mode: with the Monitor off, press and hold the Menu
button at the same time as pressing and holding the On/Off button for 3
seconds. The Monitor enters the configuration mode.
For a few seconds immediately after power up in this mode, the Systolic and
Diastolic windows display the major and minor version codes. The version codes
are numbers that represent the letters of the English alphabet, which are
designated to the currently loaded version of the monitor firmware (e.g., 1
indicates A, 2 indicates B, etc.).
At the same time, AUSC appears in the “min” (minutes display) window if the
monitor is configured with auscultatory BP. That area appears blank for a
monitor with classic DINAMAP BP configuration.
Display
Window
Major software revision Systolic
Minor software revision Diastolic
Type of BP technology min
These displays appear only during the first part of the power up sequence and
are not selectable and cannot be changed. After a moment, this version
information is cleared, and the monitor displays the 1
st
page of configuration
mode which simply displays CFG in the Systolic window.