Service manual

1-6 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Introduction: Equipment Symbols
The ProCare Monitor is designed to conform to Electromagnetic
Compatibility (EMC) standard IEC 60601-1-2 and will operate accurately in
conjunction with other medical equipment which also meets this
requirement. To avoid interference problems affecting the Monitor, do not
use the Monitor in the presence of equipment which does not conform to
these specifications.
Place the ProCare Monitor on a rigid, secure surface. Monitor must only be
used with mounting hardware, poles, and stands recommended by GE
Medical Systems Information Technologies.
The weight of the accessory basket contents should not exceed 5 lb (2.7kg).
Arrange the external AC/DC power converter, air hoses, and all cables
carefully so they do not constitute a hazard.
Verify calibration of BP parameter (temp and pulse oximeter do not require
calibration). Ensure that the display is functioning properly before operating
the ProCare Monitor.
Do not immerse the Monitor in water. If the Monitor is splashed with water or
becomes wet, wipe it immediately with a dry cloth.
Do not gas sterilize or autoclave.
Use of portable phones or other radio frequency (RF) emitting equipment
near the system may cause unexpected or adverse operation.
The equipment or system should not be used adjacent to, or stacked with,
other equipment. If adjacent or stacked use is necessary, the equipment or
system should be tested to verify normal operation in the configuration in
which it is being used.
The ProCare Monitor, when used with GE Medical Systems Information
Technologies-approved applied parts and accessories, is protected against
defibrillator damage.
NOTE: The electromagnetic compatibility profile of the ProCare Monitor may
change if accessories other than those specified for use with the
ProCare Monitor are used.
Equipment Symbols
The following symbols are associated with the ProCare Monitor. Some of the
symbols may not appear on all equipment.
NOTE: The model of the Monitor determines which symbols appear on it.
Attention, consult accompanying documents
Silence
Alarms
+ / -
Menu