Service manual

Revision D DINAMAP
®
ProCare Monitor A-9
2009381-001
Technical Specifications: Specifications
Temperature
Scale °Fahrenheit (F)
°Celsius (C)
Range
Predictive mode Max: 41.1° C; 106.0° F
Min: 35.6° C; 96.0° F
Monitor mode Max: 41.1° C; 106.0° F
Min: 26.7° C; 80.0° F
Monitor mode accuracy ± 0.1° C
± 0.2° F (when tested in a calibrated liquid bath;
meets ASTM E1112, Table 1, in range specified)
Determination time approx. 10 seconds, typical
Use only IVAC
®
probes and P850A probe covers. The size, shape, and thermal
characteristics of the probe covers can affect the performance of the
instrument. Inaccurate readings or retention problems may occur unless
IVAC
®
probes and probe covers are used. Refer to Appendix D for reorder
codes.
Factory Default Settings
Unit of measure ° F
IVAC
®
Patents
U.S. D300,728, D300,909
Pulse Rate
Factory Default Settings
Pulse Rate HIGH: 150
LOW: 50
Refer to individual SpO
2
and BP sections.