Technical Specifications: Specifications
‡The Masimo
SET
®
SpO
2
parameter has been validated for low perfusion accuracy in bench top testing
against a Biotek Index 2 stimulator and Masimo’s simulator with signal strengths of greater than 0.02% and
a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or
minus, one standard deviation.Plus or minus one standard deviation encompasses 68% of the population.
Masimo
®
Sensor Accuracy
Sensor Model SpO2 Range
70% - 100%
LNOP
LNOP-ADT ± 2 digits
LNOP-ADT Long ± 2 digits
LNOP-PDT ± 2 digits
LNOP-NEO ± 3 digits
LNOP-NEO PT ± 3 digits
LNOP-DCI (reusable) ± 2 digits
LNOP-DCSC (reusable) ± 2 digits
LNOP-DCIP (reusable) ± 2 digits
NRI25 (reusable) ± 2 digits
LNOP-DC195 (reusable) ± 2 digits
Resolution
Saturation (% SpO
2
)1%
Pulse Rate (bpm) 1
Low Perfusion Performance
>0.02% Pulse Amplitude Saturation (% SpO
2
) ±2 digits
and % Transmission >5% Pulse Rate ±3 digits
Interfering Substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is
approximately equal to the amount of carboxyhemoglobin present. Dyes, or any
substance containing dyes, that change usual arterial pigmentation may cause
erroneous readings.
Sensor Light Source
Wavelength Infrared: 905 nm (nominal)
Red: 660 nm (nominal)
Power Dissipation Infrared: 22.5 mW (max)
Red: 27.5 mW (max)
Factory Default Settings
SpO
2
(%) HIGH: 100
SpO
2
(%) LOW: 90
Sensitivity Mode 2 (for low perfusion-Default)
Averaging Time 12 seconds
FastSAT Mode 0 (for Off)
Masimo Patents
5,482,036; 5,490,505; 5,632,272; 5,685,299; 5,758,644; 5,769,785; 6,002,952;
6,036,642; 6,067,462; 6,206,830; 6,157,850, and international equivalents.