Service manual

A-6 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Technical Specifications: Specifications
**Applicability:
OxiMax
®
MAX-A, MAX-AL, MAX-P, MAX-I, and MAX-N sensors.
NELLCOR
®
Sensor Accuracy
NOTE: All NELLCOR
®
sensors must be used with the NELLCOR
®
DOC-10 cable;
the SCP-10 cable is not compatible with the ProCare Monitor.
Sensor Model
SpO
2
Range
70% - 100%
OxiMax
®
MAX-A*, MAX-AL* ±2 digits
MAX-N*
(Adult) ±2 digits
MAX-N*
(Neonate) ±3 digits
MAX-P* ±2 digits
MAX-I* ±2 digits
MAX-R*
±3.5 digits
OxiCliq
®
OxiCliq A ±2.5 digits
OxiCliq P ±2.5 digits
OxiCliq N
(Adult) ±2.5 digits
OxiCliq N
(Neonate) ±3.5 digits
OxiCliq I ±2.5 digits
Reusable Sensor Models
D-YS (Infant to Adult) ±3 digits
D-YS (Neonate) ±4 digits
D-YS with D-YSE ±3.5 digits
D-YS with D-YSPD ±3.5 digits
DS-100A ±3 digits
OXI-A/N (Adult) ±3 digits
OXI-A/N (Neonate) ±4 digits
OXI-P/I ±3 digits
* The accuracy specification under motion conditions is ±3. For a definition of motion, contact
NELLCOR
®
Technical Services or your local representative.
The MAX-N and the OxiCliq N were tested on patients >40 kg.
The accuracy specification has been determined between saturations of 80%-100%.
Neonatal Sensor Accuracy
When sensors are used on neonatal subjects as recommended, the specified
accuracy range is increased by ±1 digit, to account for the theoretical effect on
oximeter measurements of fetal hemoglobin in neonatal blood. For example,
MAX-N accuracy on neonates is ±3 digits, rather than ±2 digits.
Sensor Light Source
Wavelength Infrared: 890 nm (nominal)
Red: 660 nm (nominal)