Service manual
A-4 DINAMAP
®
ProCare Monitor Revision D
2009381-001
Technical Specifications: Specifications
Environmental
Operating Temperature + 5 °C to + 40 °C
(+ 41 °F to + 104 °F)
Operating Atmospheric
Pressure 700 hPa to 1060 hPa
Storage Temperature – 20 °C to + 50 °C
(– 4 °F to + 122 °F)
Storage/Transportation
Atmospheric Pressure 500 hPa to 1060 hPa
Humidity Range 5% to 95% noncondensing
Radio Frequency Complies with IEC Publication 60601-1-2 Medical
Electrical Equipment, Electromagnetic
Compatibility Requirements and Tests and CISPR
11 (Group 1, Class B) for radiated and conducted
emissions.
BP
Cuff Pressure Range 0 to 290 mmHg (adult/ped)
(Normal operating range) 0 to 140 mmHg (neonate)
Blood Pressure Accuracy Meets ANSI/AAMI standard SP-10 (mean error ≤5
mmHg, standard deviation ≤8 mmHg).
Auscultatory: BHS protocol Grade A/A; ESH
protocol requirements passed
Maximum Determination 120 s (adult/ped)
Time 85 s (neonate)
Overpressure Cutoff 300 to 330 mmHg (adult/ped)
150 to 165 mmHg (neonate)
Pulse Rate Range
30 to 200 beats/min (adult/ped)
30 to 220 beats/min (neonate)
Pulse Rate Accuracy ± 3.5%
Factory Default Settings
Systolic Limits (mmHg) HIGH: 200
LOW: 80
Diastolic Limits (mmHg) HIGH: 120
LOW: 30
Inflation Pressure* 160 mmHg (adult/ped)
110 mmHg (neonate)
Cycle button 15
*Default inflation pressure can be changed in Config Mode. Refer to the Getting Started section of
operation manual.