5.6.1 Calibration Adapter Plug Doesn’t Fit
Use a slight twisting motion as you gently push the calibration adapter plug (p/n: CET-7000-
CAP) into the sensor opening. If it is hard to insert, moisten the O-ring seal slightly then try
re-inserting it.
5.6.2 Zero Fault
If the zeroing process fails, the cGas Detector will show a Zero fault. This will happen if the
ambient gas readings are at an unacceptable level due to not enough clean air - there is
enough residual target gas in the environment or other gases that are interfering with the
sensor seeing oxygen.
5.6.3 Span Fault
If the cGas shows a Span fault after trying to calibrate, it is possible that the sensor is dead
or the device isn’t seeing the gas (cylinder is empty, regulator or adapter plug is not attached
properly, tube is blocked, kinked, etc.).
Span Fault
197 AD 1
0s
Check all connections and possible interferences and try a complete calibration procedure
from the beginning again to see if this corrects the fault. If a second calibration does not
resolve the fault then the sensor needs to be replaced. Over time, a sensor degrades and
when it has gone beyond an acceptable level, it has reached its end of life and will no longer
pass a calibration. If replacing the sensor does not correct the fault, please contact Technical
Service at help@cetci.com.
5.6.4 Zero Override
If the gas level (possible residual gas) is too high, but still within the override range, the
display will indicate that an override is required.
Override?
-?????
N
>Y
To override the value use the UP button to select Y and press ENTER. To keep the original
zero value, leave the indicator on N and press ENTER.
5.6.5 Span Override
During the Span calibration, readings are taken and from the results sensitivity is calculated
and compared to the original sensitivity of the sensor at the time of installation. If this
sensitivity is below the override range, but above the fault limit, the display will indicate that
an override is required. To override the value use the UP button to select Y and press ENTER.
To keep the original value, leave the indicator on N and press ENTER.
Override?
1953 AD 19
N
>Y
6 SENSOR REPLACEMENT
p/n: VLT-NH3-SEN
The catalytic ammonia sensor has a lifespan of approximately 2 to 3 years under normal
air conditions. The sensor may need replacing sooner if is is poisoned by exposure to a very
high concentration of ammonia.