User's Manual

16
CPR BAND (Model: CREDO-CB)
CPR Mode
1) Compression Depth Feedback
The LED lights according to the correct compression depth set in the
[Setting] tab. The maximum value is 10 cm and the minimum value is 1
cm of the correct compression depth you can set. The factory-set values
of correct range is 4.5 cm ~ 6.0 cm.
* DC: Depth of Compression
* M: Maximum depth of setting
* m: Minimum depth of setting
* A1: m
󰇛
M m
󰇜
* A2: m
󰇛
M m
󰇜
Compression Depth
DC < m Lower 1 orange LED lights up
m ≤ DC < A1 1 blue LED lights up
A1 ≤ DC < A2 2 blue LEDs light up
A2 ≤ DC < M 3 blue LEDs light up
M DC 3 blue LEDs and upper Orange LED light
up