Instruction manual
accQcomm Interface Module
Section 2 Installation and Operation
2-15
2.7 Alarm and Output
Parameters
Table 2-2 contains the parameters shown in the pull-down menus
for selecting alarm and output functions.
Table 2-2 Alarm and Output Parameters
Name Function Alarm
4-20
Output
Flow (Q Main) Flow rate calculated from cross-sectional velocities
Q Manning Flow rate calculated with Manning formula, for flow depths that fall below
the velocity profiling range
Smoothed flow Flow data with smoothing applied.
Totalized flow Total flow, including both positive and negative flow.
Velocity (Raw Velocity) Average of individual point velocities from all velocity beams.
Avg. Velocity Flow divided by cross-sectional area
Smoothed Velocity Velocity data with smoothing applied.
Main Depth (Level) This is the depth used for flow calculation, after comparing the readings
from the primary depth sensor with the readings from the secondary depth
sensor.
Ult. Depth Liquid level from ultrasonic depth sensor
Sec. Depth Liquid level from secondary depth sensor
Smoothed Depth Level data with smoothing applied.
Sensor Temperature Temperature in stream
Voltage Input voltage
Built-in Test A code ‘0’ indicates that there are no problems detected.
Average Flow Average flow reading of all active instruments
Average Velocity Average velocity reading of all active instruments
Average Depth Average liquid level reading of all active instruments
Data Timeouts: Number of timeouts on counter
Average Temperature Average temperature reading of all active instruments
Summed Flow Summed flow reading of all active instruments
Bad Flow Count Number of consecutive readings with a bad value
Bad Velocity Count Number of consecutive readings with a bad value
Bad Depth Count Number of consecutive readings with a bad value