Instruction manual

accQcomm Interface Module
Section 2 Installation and Operation
2-12
Note
For the accQmin flow meter, the accQcomm displays and
transmits secondary level data at 10X the actual value when
using Standard Data format. This should be considered when
setting up alarm conditions and analog outputs. This will not
occur with an accQmin using Reduced Data format.
Testing Analog Outputs To verify that the analog outputs are functioning properly.
1. Attach the leads of a current meter to the output circuit to
be tested.
2. Click Test Analog Outputs to open the test dialog.
Figure 2-10 Analog output test dialog
3. Select a COM port and baud rate for communication.
4. Select the channel and enter a current value in mA.
5. Click Test Output.
The measured current value should match the set value,
until or unless changed by another event or action.