Instruction manual

accQcomm Interface Module
Section 2 Installation and Operation
2-9
When you highlight the desired alarm number in the Current
Alarms window and click Set Alarm Condition, the alarm
appears beside its number.
To remove an alarm condition, highlight the alarm and click
Clear Selected.
Note
For the accQmin flow meter, the accQcomm displays and
transmits secondary level data at 10X the actual value when
using Standard Data format. This should be considered when
setting up alarm conditions and analog outputs. This will not
occur with an accQmin using Reduced Data format.
Sampler Pacing The alarm function can signal a connected sampler to take a
sample every time a specified volume of flow is reached.
In the Sampler Pulse box, select the instrument and enter the
volume at which to send the pulse. When you highlight the
desired alarm number in the Current Alarms window and click
Set Pulse Volume, the setting appears as an alarm.
A separate sampler pulse output can be set for each connected
instrument, if needed.
2.5 Outputs The accQcomm can make the flow meter data available in
several formats. Data output is configured on the Output tab.
From the Output Type pull-down menu, select either MODBUS
or Analog (4-20mA).
2.5.1 MODBUS Selecting MODBUS from the pull-down menu activates the
MODBUS settings at the bottom of the screen. Select ASCII or
RTU protocol from the pull-down menu, as well as appropriate
serial data settings.
Internal units of measure for MODBUS cannot be changed.
Parameter values will always be output as follows:
Note
For the accQmin flow meter, the accQcomm displays and
transmits secondary level data at 10X the actual value when
using Standard Data format. This should be considered when
setting up alarm conditions and analog outputs. This will not
occur with an accQmin using Reduced Data format.
Flow cc/sec
Area cm
2
Te mp ° C
Velocity mm/sec
Depth mm
Volume m
3