User manual

User Manual for Multi-parameter Patient Monitor
61
The high and low limits setting range:
Type
Parameter
Adult Infant Neonate
Heart Rate
High Limit
1~300
bpm
(1~350) bpm
(1~350) bpm
Low Limit
0~299
bpm
(0~349) bpm
(0~349) bpm
Respiration Rate
High Limit
1~120
rpm
(1~150)rpm
(1~150)rpm
Low Limit
0~119
rpm
(0~149)rpm
(0~149)rpm
Temperature
High Limit
0.1~60.0
)℃
(0.1~60.0)
(0.1~60.0)
Low Limit
0 ~49.9
)℃
(0 ~49.9)
(0 ~49.9)
NIBP
Systolic
High Limit
31~270
mmHg
(31~200)mmHg
(31~135)mmHg
Low Limit
30~269
mmHg
(30~199)mmHg
(30~134)mmHg
Diastolic
High Limit
11~232
mmHg
(11~150)mmHg
(11~100)mmHg
Low Limit
10~231
mmHg
(10~149)mmHg
(10~99)mmHg
MAP
High Limit
21~242
mmHg
21~165
mmHg
(21~110)mmHg
Low Limit
20~241
mmHg
(20~164)mmHg
(20~109)mmHg
SpO
2
High Limit
1~100
)%
(1~100)
(1~100)
Low Limit
0~99
)%
(0~99)
(0~99)
Pulse Rate
High Limit
1~300
bpm
(1~350)bpm
(1~350)bpm
Low Limit
0~299
bpm
(0~349)bpm
(0~349)bpm
ST Segment
High Limit
-2.49mV~+2.49mV
-2.49mV~+2.49mV
-2.49mV~+2.49mV
Low Limit
-2.49mV~+2.49mV
-2.49mV~+2.49mV
-2.49mV~+2.49mV
Temperature
Difference
0.0~5.0)℃ (0.0~5.0) (0.0~5.0)
Alarm high and low limits: when the parameter value is higher than the high limit or lower than the low limit, the
system will alarm.
In the monitoring screen, press the Alarm key to activate the alarm timer. The time shows up on the high limit left
corner of the screen. When the alarm timer is activated, the system will activate the alarm function when the
specified time has passed.
Alarm timer has 4 options: 2 min, 5 min, 10 min, 20 min, or during alarm.