User manual

User Manual for Multi-parameter Patient Monitor
60
Default Alarming Values of All Parameters and Setup Range
The default alarming value:
Type
Parameter
Adult Infant Neonate
Heart Rate
High Limit
180bpm
200bpm
220bpm
Low Limit
40bpm
50bpm
50bpm
Respiration
Rate
High Limit
40rpm
50rpm
60rpm
Low Limit
10rpm
10rpm
10rpm
Temperature
High Limit
39
39
39
Low Limit
35
35
35
NIBP
Systolic
High Limit
180mmHg
130mmHg
110mmHg
Low Limit
60mmHg
50mmHg
50mmHg
Diastolic
High Limit
120mmHg
90mmHg
90mmHg
Low Limit
50mmHg
40mmHg
30mmHg
MAP
High Limit
160mmHg
110mmHg
100mmHg
Low Limit
50mmHg
40mmHg
30mmHg
SpO
2
High Limit
100
100
100
Low Limit
90 85 85
Pulse Rate
High Limit
180bpm
200bpm
220bpm
Low Limit
40bpm
50bpm
50bpm
ST Segment
High Limit
+1.00mV
+1.00mV
+1.00mV
Low Limit
-1.00mV
-1.00mV
-1.00mV
Temperature
Difference
Range
2 2 2