User Manual for Multi-parameter Patient Monitor
56
In order to avoid damaging the battery, do NOT use other power supply device to charge the
battery;
After battery ageing phenomenon occurring, do NOT throw the battery into fire to avoid
explosion risk.
Do not hit or strike it with force;
Do not use this battery on other devices;
Do not use this battery below -10
or above 40
;
In order to maintain battery supply time and prolong battery lifetime, please use the battery once
a month and do not charge it until it is used up each time.
9.2 Cleaning, Sterilization and Disinfection
Switch off the monitor and disconnect the power cable before cleaning.
Kept the monitor from dust.
It is recommended to clean the outer shell and screen of the monitor to keep it clean. Only non-corrosive
cleanser such as clear water is permitted.
Wipe the surface of the monitor and transducers with an alcohol impregnated wipe, and dry it with dry
and clean wipe or simply air-dry.
The monitor can be disinfected and sterilized, please clean the monitor first.
Do not let the liquid cleanser flow into the connector jack of the monitor to avoid damage.
Clean the exterior of the connector only.
Dilute the cleanser.
Do not use scrub materials.
Do not let any liquid flow into the shell or any parts of the monitor.
Do not let the cleanser and disinfectant stay on its surface.
Do not perform high pressure sterilization to the monitor.
Do not put any parts of the monitor in the liquid.
Do not pour the disinfector on its surface while sterilization.
9.3 Cleaning, Sterilization and Disinfection of Accessories
It is recommended to clean the accessories (including sensor, leads and plugs) with a piece of gauze which has
been soaked in 75% Alcohol or 70% Ispropanol.
Do not use damaged accessories.
Accessories can not be entirely immerged into water, liquor or cleanser.
Do not use radial, steam or epoxyethane to disinfect accessories.