User manual
User Manual for Multi-parameter Patient Monitor
51
6. Subject with hyperadiposis;
In addition, statistics show that 37% people report blood pressure difference of no less than 0.80kPa(6mmHg)
between the left and right arms, and 13% people report difference of no less than 1.47kPa (11mmHg).
Note: Some practitioners may report big discreteness or abnormal value of the blood pressure measures
when the oscillating method is used. As a matter of fact, the so-called “big discreteness” must be a
term in the sense of statistical significance of mass data. Abnormal data may be observed in some
individual cases. It is normal in the scientific experiments. It may be caused by an apparent reason,
or by an unknown factor in some cases. Such individual doubtful experimental data may be
identified and eliminated using the special statistical technique. It is not a part of this manual. The
practitioner may eliminate the apparently unreasonable data according to the experience.
Operation Introduction:
1. Take a measurement in manual mode:
Enter into the screen of NIBP setting, select “Mode” option and set it as “MANU”, and then
press the NIBP key on the front panel to start measure. If press the NIBP key again, the
measurement will be stopped.
During the automatic measurement interval when no NIBP measurement is taken, press the NIBP
key, a measurement in manual mode will be taken. If at this time press the NIBP key again, the
manual mode will be stopped and continue the automatic mode.
2. Take a measurement in automatic mode:
Enter into the screen of NIBP setting, select “Cycle” option and select time interval according needs,
then press the NIBP key on the front panel to start the automatic measurement at a certain interval.
3. Stop automatic measurement
In the procedure of automatic measurement, press the NIBP key at any time, the measurement will be
stopped.
4. STAT measurement
Enter into the screen of NIBP settings, select Cycle option and set as STAT, the STAT measurement will
be taken. This procedure will last for 5 minutes.
5. Stop STAT measurement on the halfway
In the procedure of STAT measurement, press the NIBP key at any time, the measurement will be stopped.
7.4 SpO2/Pulse Monitor
7.4.1 SpO2 Monitor
This monitor measures the pulse oxygen saturation (SpO
2
) and pulse by means of the radiograph of infrared light
and the red light emitted by LED through body’s peripheral areas (such as fingers), whereby the photoelectric
detecting circuits will analyze the absorptivity of the oxyhemoglobin and reduced hemoglobin respectively, and
give the photoabsorption rates before and after pulsation. Using the measure of photoabsorption change due to
pulsatory arterial blood flow caused by PLETH waveform, the SpO
2
can be obtained.
7.4.2 Points to be Noted in SpO2 and Pulse Measuring
1. The fingers should be properly placed (see the attached illustration of this instruction manual), or else it
may cause inaccurate measurement result.