User Manual for Multi-parameter Patient Monitor
42
5.4 TEMP Monitoring
1. TEMP measuring range: 25.0 ~45.0
2. TEMP measuring accuracy: ±0.2
3. TEMP response time: ≤150s
5.5 NIBP Monitoring
1. Measuring method: Automatic measurement by intelligent oscillating method.
2. Cuff inflation time: <30 seconds (typical adult cuff)
3. Air release time during measurement: < 90 seconds on the average
4. Air release time while the measurement is canceled: <2 seconds (typical adult cuff)
5. Initial cuff inflation pressure: Adult and infant mode: 180mmHg, Neonate mode: 120mmHg.
6. Overpressure protection limit: Adult and infant mode: 300 mmHg, Neonate mode: 150 mmHg.
7. Blood pressure measuring range:
Systolic pressure: Adult and infant: 25mmH~265mmHg; Neonate: 25mmHg~135mmHg
Diastolic pressure: Adult and infant: 12mmHg~220mmHg; Neonate: 12mmHg~110mmHg
Mean arterial pressure: Adult and infant: 18mmHg~245mmHg; Neonate: 18mmHg~120 mmHg
8. Measurement accuracy Systolic ±10%
Diastolic ±10%
MAP ±10%
9. Measurement mode: Manual, Auto, STAT
Automatic measuring intervals:1~240min; interval tolerance: <10s
5.6 SpO
2
& Pulse Rate Monitoring
1. Transducer: dual-wave length LED
2. SpO
2
measuring range: 35%~100%
3. Low perfusion capability: as low as 0.4%
4. SpO
2
display accuracy:
(1) 75%~99% ±2%
(2) 50%~75% ±3%
5. Pulse rate measuring range: 30bpm~250bpm
6. Pulse rate measuring accuracy: ±2bpm
5.7 Recorded Data
1. Sensitivity selection tolerance: ≤5%
2. Recording speed: 25mm/s
3. Recording speed accuracy: ≤5%