Chapter 5 Technical Specifications
5.1 Cerebral State Monitoring
1. EEG sensitivity: ±400μV
2. Noise level: <2μVp-p, <0.4μV RMS, 1~250Hz
3. CMR: >140dB
4. Input impedance>50 Mohm
5. Sample rate: 2000 samples/sec. (14 bits equivalent)
6. CSI and update: 0-100. Filter 6-42 Hz, 1sec. update
7. EMG: 0-100 logarithmic. Filter 75-85 Hz, 1 sec. update
8. BS%: 0-100%. Filter 2-42 Hz, 1sec. update
9. Digital output: Wireless to R232 link (ISM 2.4 GHz)
10. Wireless range: Up to 10 meters
5.2 ECG Monitoring
1. Heart rate measurement: ±0.4mVp~±3mVp
2. Heart rate display range: 20bpm~250bpm
3. Heart rate display accuracy: ±1% or ±2bpm, whichever is greater
4. Heart rate alarm delay time: ≤12s
5. Sensitivity selection: Auto, ×1/2, ×1, ×2, ×4 tolerance: ≤5%
6. Sweeping speed: 12.5mm/s, 25mm/s, 50mm/s tolerance: ≤10%
7. ECG noise level: ≤30μ
V
P-P
8. ECG input loop current: ≤0.1μA
9. Input impedance: ≥5MΩ
10. Common-mode rejection ratio (CMRR): ≥80dB
11. Time constant:
Monitoring mode≥0.3s Diagnostic mode≥3.2s
12. Frequency response:
Monitoring mode: 0.5Hz ~ 25Hz Diagnostic modes: 0.05Hz~75Hz
5.3 RESP Monitoring
1. RESP monitoring range: 0rpm~120rpm
2. RESP monitoring accuracy: ±10%
3. RESP alarm range: 0bpm~100bpm. Alarm tolerance: ±10%