ID: or Patient ID. To enter patient ID, choose the patient ID field using the Navigation Knob. Press it to
enter the text entry box. Rotate the knob to choose the letter and press the knob to enter the letter. To delete
the letter, move the cursor to the letter to be deleted and rotate the knob to enter spaces (after the H). Use
the spaces to replace the letters. After finishing entering the patient ID, choose Exit button and press the
knob to exit the text entry. The patient ID is the unique identifier for the patient. When the patient ID
changes, the system considers the patient has changed.
Name
: Enter the patient’s name.
Bed:
Enter the patient’s bed number.
Sex
: Choose between M or F for male and female.
Age
: Choose the age field and use the rotate knob to select an age.
Save
: The operator can choose how much of the data that needs to be saved. The unit is hour. Once the time
is chosen, the system starts to save data from the current time. If OFF is selected, it means that the data will
not be saved. The system will determine the time range according to the disk space available. If no disk
space available, SAVE will be displayed as OFF. When the user intends to save the current ECG waveform
permanently, please delete the history files.
☞ The printing is determined by the option of CSM item on Print Setup Screen if you press the Print Key, refer
to Chapter 4.9.2 Printer Parameter Settings for details.
4.12 Event List
In the Mode Selection screen, move the cursor to the “ Event”, and press the Navigation Knob to enter event list
screen, shown in Figure 4.39.
Figure 4.39 Event List
In event list screen, the event time, number and type of the record during anaesthesia can be viewed.
Event type: induction,intubation,maintenance,surgery,injection,note, end maintenance,movement etc.
☞ Press the Print Key to print Event List.