User manual
User Manual for Multi-parameter Patient Monitor
38
4.9.6 SpO
2
Parameter Settings
SpO
2
Hi/Lo:
high and low limits of SpO
2
alarm
Pulse Hi/Lo: high and low limits of pulse rate alarm
Figure 4.35 Respiration Parameter Settings
Gain: Respiration amplification/gain, 4 options, x1/2, x1, x2, x4, the default is x1
Speed: Respiration display speed. 2 options 6.25mm/s and 12.5 mm/s. The factory setting is 12.5 mm/s
Apnea: The apnea alarm time (in second). When the patient stop breathing for the time longer than the
set period, the Respiration display channel display warning “Apnea xxx second”.
Type: Respiration impedance
RR Hi: High limit of respiration rate alarm
Lo: Low limit of respiration rate alarm
Figure 4.36 CSM Parameter Settings
4.9.7 CSM Parameter Settings
CSI Hi
:
High limit of CSI alarm
Lo
:
Low limit of CSI alarm
Scope
:
select the scope of CSM waveform , the unit is
μ V.