For setting temperature alarm range, if the temperature difference is higher than the setting value,
the system will alarm.
Figure 4.33 NIBP Parameter Settings
4.9.5 NIBP Parameter Settings
Unit:
The unit of NIBP. The unit mmHg or kPa can be selected.
Mode:
The cuff inflation mode, manual or automatic. The factory default is manual. The operator
needs to press the NIBP button to perform NIBP measurement. If the Auto mode is chosen, the operator
needs to set an interval cycle as well.
Cycle:
The inflation interval when the NIBP measurement is set to Auto. The options are STAT, 1 min,
2 min…240 min. Press NIBP key and the system begins to count down. It takes blood pressure
measurement automatically after finishing counting down If STAT is selected, press the NIBP Key to
take 5 minutes’ measurement.
WARNING: STAT can only be used for Adult. Using this mode to neonate can cause
serious injury.
SYS Hi/Lo:
high and low limits of systolic pressure alarm
DIA Hi/Lo:
high and low limits of diastolic pressure alarm
MAP Hi/Lo:
high and low limits of mean arterial pressure alarm
Unit:
the pressure unit. mmHg or kPa. The factory default is mmHg.
Figure 4.34 SpO
2
Parameter Settings