User manual
User Manual for Multi-parameter Patient Monitor
36
DIA: Diagnosis. No filtering, represent the true ECG with out filtering.
OPE: Operation. Deep filtering, filtering out strong interference.
The factory default is MON.
BtSnd: Heart beat sound. The synchronous heart beat sound during monitoring. The factory default
is ON.
1mV: Generating the 1mV signal. This signal is used to test the function of the machine. It is not
used during normal operation. Factory default is OFF.
ST Hi: high limit of ST segment alarm
Lo: low limit of ST segment alarm
50Hz: 50 Hz frequency filter. The factory default is ON.
Pace
:
Pace detecting switch. The default is “OFF”. When the option is on, the mark will be made in
the first ECG channel of any screen if pace is detected, as shown below.
Figure 4.31 Heart Beat Detection Screen
☞ Limits setup: Move the gray cursor to the high or low limits of the alarm settings, and press the
(Alarm) key to turn ON or OFF the alarm for the setting. Yellow color shows ON status, and
gray color show the OFF status.
ST High
:
ST segment high limit alarm
Low
:
ST segment low limit alarm
Figure 4.32 Temperature Parameter Settings
4.9.4 Temperature Parameter Settings
TEMP1 Hi
:
High limit of temperature 1 alarm
Lo
:
Low limit of temperature 1 alarm
TEMP2 Hi:High limit of temperature 2 alarm
Lo:Low limit of temperature 2 alarm
Unit:Temperature unit. The default is °C (Celsius) and it can be set to °F (Fahrenheit).