User manual
User Manual for Multi-parameter Patient Monitor
35
Printer:
Printer switch. If the printer option is “ON”, the printer will be available. This parameter
does not have default, but the choice can be stored.
Timer: It is the switch of the automatic print timer. When the set is “ON”, 1 minute to 240 minutes
(the intervals at which the system will print the information automatically) adjustable. When the time
setting is reached, the system will automatically print waveform information. The default is “OFF”.
Wav2: When the Printer option is “ON”, it will be available. The first printed waveform is Lead II
ECG waveform, and the second printed waveform is optional. The SpO
2
, respiration, and Lead I, III,
AVR, AVL, AVF, V ECG waveforms can be selected. The default is “SpO
2
”
.
VPC: When the option is “ON”, the printer will be triggered on when arrhythmia appear, and record
waveform data automatically. The default is “OFF”.
CSM: Printing option in the main screen. When the option is “OFF”, press the Print key in the main
screen, a channel of ECG waveforms and the channel of “Wav2” will be printed. When the option is
“PARA”, CSM trend will be printed; when the option is “EEG”, EEG waveform will be printed.
Note: When printing CSM trend graph, different parameters use different types of line to show the
waveform. CSI: medium-thick solid line; BS: thick dotted line; EMG: Single solid line; SQI: Single
dotted line
.
Figure 4.30 ECG Parameter Settings
4.9.3 ECG Parameter Settings
Lead: Can choose from I, II, III, AVR, AVL, AVF, V (V1-V6). The default is I.
Gain: The ECG gain, 5 options x1/2, x1, x2, x4 and Auto. Auto is for automatic gain control. The
factory default is x1
HR Hi: high limit of heart rate alarm
Lo: low limit of heart rate alarm
The adjustable range and the factory default value can be found in chapter 5.
Speed: ECG display speed. 3 options: 12.5, 25, 50 mm/s. The factory default is 25 mm/s
Mode: ECG filter mode. Three options: MON, DIA, and OPE
MON: monitoring mode. Moderate filtering that can filter out interference and present good
ECG waves.