The Default button is used to resume default vales of parameter, but the Document information and its data will
not be changed. We will cover the functions of each button below.
On any parameter setting screen, press print key to print parameter settings.
4.9.1 System Parameter Settings
Type
: The subject being monitored, this can be selected among Adult, Infant and Neonate.
Adult
: the subject is adult.
Infant
: the subject is pediatric patient.
Neonate
: the subject is neonate.
The default is “Adult”
When changing the patient type, the system will perform the alarm settings, NIBP settings
initializations. Please pay special attention to the patient type before starting the monitoring. It is
forbidden to use Adult type on the infant patient. Doing so can cause serious injury.
Mode:
Monitor mode selection. The Real Time shows the real time waveform, i.e. normal
monitoring state. The DEMO shows the demo waveforms. In the demo state, all the signals and data
are generated from the patient monitor for demo and testing purpose. The default is “Real Time”
Lang:
the current language used, which can be selected by the user. There is no default for this
setting. However, the setting can be saved.
Fill:
When the fill setting is ON, the display fills the volume for the SpO
2
and Respiration. When it is
OFF, the system displays the line graph. The default is OFF.
Frze:
Pressed the key to freeze the selected waveform. The options are “All” and “ECG”. When
ECG is selected, the system only freezes the ECG waveform. When “All” is selected, the system
freezes all the waveforms including ECG, SpO
2
, and Respiration. The factory default is “ECG”
Disp2:
The
screen can display the observe screen or seven ECG waveforms on the same screen.
VOL:
The sound volume. The maximum volume is 7 and minimum is 0, i.e. no sound. The default is
5.
Key:
If the setting is ON, the press of the button will generate a keystroke sound. The factory is ON.
BP source:
Blood pressure value display option, the default is “NIBP”.(not adjustable)
4.9.2 Printer Parameter Settings
Figure 4.29 Printer Parameter Settings